高纯氧化镥中痕量铁的富集及测定研究

Enrichment and determination of trace iron in high purity lutetium oxide

  • 摘要: 采用吡咯二硫代氨基甲酸铵(APDC)沉淀富集高纯氧化镥中痕量的铁,再将沉淀溶于甲基异 丁基酮(MIBK)溶液中,分离富集后的溶液用石墨原子吸收法直接测定.当盐酸酸度控制在0.4 mol/L,加 2 %APDC溶液2 mL,沉淀放置时间为15 min,选取石墨炉原子吸收光谱仪的灰化温度1 100℃,原子 化温度2 300 ℃时,铁的测定效果最佳.铅、镍、铬、钴、铜等基体元素对铁的测定基本无干扰.该方法 具有较好的精密度和准确度,相对标准偏差为4.5 %,加标回收率在94.0 %~108.0 %之间.

     

    Abstract: Ammonium pyrrolidine dithiocarboxylic acid (APDC) was used to subside and gather trace iron of high purity lutetium oxide, and the precipitate of which was dissolved in methyl isobutyl ketone (MIBK). After separation and enrichment, the solution was determined by GF-AAS. When the acidity of liquid is 0.4 mol/L, the volume of 2 % APDC is 2 mL, settling time is 15 min, ashing temperature is 1 100 ℃, and atomization temperature is 2 300 ℃, determination result of Fe is the optimum. Matrix elements such as Pb, Ni, Cr, Co and Cu have no interference on the determination of Fe. This method is comparatively accurate and precise with relative standard deviation of 4.5 %, and the recovery rate within the range of 94.0 %~108.0 %.

     

/

返回文章
返回